Skip to content

Public FDA record

Z-0606-2011: Beaver-Visitec International (Beaver-Visitec) BD 1% OVD Sodium Hyaluronate, 10 mg/ml 1.0 mL. Sterile. Rx

Recall record Z-0606-2011 for Becton Dickinson and Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0606-2011
Company in FDA record
Becton Dickinson and Company
FDA classification date
Dec 16, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Finger flange on the BD 1% OVD syringe can disengage during cataract surgery and and may lead to injury.

Source facts

Action
Beaver Visitec issued an Urgent: Field Safety Notice letter, dated 10/4/10 to direct accounts, identifying the affected product and informing users of the potential for the finger grip to disengage during use. The firm offers solutions as follows: Allow the finger grip to remain attached to the syringe and continue to use the product. Position the open side of the finger grip towards the palm to avoid detachment of the grip from the syringe. Remove the finger grip and continue to use the product. Return the unused product to Beaver-Visitec for reimbursement. Users were requested to complete the Customer Response Form. Questions can be directed to: Inside Sales at 781-906-7952.
Initiated
2010-10-04
Root cause
Component design/selection
Terminated
2012-02-16
Distribution
Nationwide Distribution
Recall number
Z-0606-2011
Recall status
Terminated
CDRH record ID
95400
FDA classified
2010-12-16
Recall event number
57104

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗