Open original FDA record ↗Recall record
- FDA identifier
- Z-0606-2011
- Company in FDA record
- Becton Dickinson and Company
- FDA classification date
- Dec 16, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Finger flange on the BD 1% OVD syringe can disengage during cataract surgery and and may lead to injury.
Source facts
- Action
- Beaver Visitec issued an Urgent: Field Safety Notice letter, dated 10/4/10 to direct accounts, identifying the affected product and informing users of the potential for the finger grip to disengage during use. The firm offers solutions as follows: Allow the finger grip to remain attached to the syringe and continue to use the product. Position the open side of the finger grip towards the palm to avoid detachment of the grip from the syringe. Remove the finger grip and continue to use the product. Return the unused product to Beaver-Visitec for reimbursement. Users were requested to complete the Customer Response Form. Questions can be directed to: Inside Sales at 781-906-7952.
- Initiated
- 2010-10-04
- Root cause
- Component design/selection
- Terminated
- 2012-02-16
- Distribution
- Nationwide Distribution
- Recall number
- Z-0606-2011
- Recall status
- Terminated
- CDRH record ID
- 95400
- FDA classified
- 2010-12-16
- Recall event number
- 57104