Possible false-negative test results: The Sample A Diluent Lot # 50768 assigned to Is-Measles IgG Lot #'s 41108 and 41108B has higher conductivity than normal, which can sporadically interfere with the MAGO Plus/Aptus sensing of and picking up of samples. As a result, if the sample is not picked up/processed, it could lead to the possibility of obtaining false negative results in some cases.
Source facts
Action
Notification of all domestic customers via Certified Mail. Notification of the one international customer via Federal Express.
Initiated
2008-10-10
Root cause
Other
Terminated
2010-05-12
Distribution
Worldwide Distribution --- including USA and country of England.
Recall number
Z-0607-2009
Recall status
Terminated
CDRH record ID
74884
FDA classified
2009-01-07
Recall event number
50162