Open original FDA record ↗Recall record
- FDA identifier
- Z-0607-2014
- Company in FDA record
- Baxter Healthcare Corp.
- FDA classification date
- Dec 29, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Following a retrospective review for Baxter Daytona Complaint CR-954, it was identified that the IntelliFill i.v. Pharmacy Compounding Device (PCD) (product code: IFIV) has the potential to continue to run and fill syringes with the HEPA filtration disabled.
Source facts
- Action
- A Service Bulletin was sent to all IntelliFill i. v. customers February 22, 20 I 0 by USPS mail. The Service Bulletin informs customers that IntelliFill i.v. operators are to be trained to listen for the running HEPA filtration fan motor and observe/feel for airflow coming from the automation deck. The field service engineer dispatched to the site replaced the damaged sensor on 11-Feb-2010 and a Service Bulletin was sent to all customers regarding the function and use of the HEPA filter on 22-Feb-2010. A new HEPA sensor was designed and installed into all current IntelliFill i.v. devices between 21-July-2011 to 9-Feb-2012.
- Initiated
- 2013-11-08
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-03-23
- Distribution
- Nationwide Distribution including the states of CA, CO, CT, FL, IL, LA, MA, MD, MI, NC, NE, NY, OH, OR, PA, TN, TX, and WV.
- Recall number
- Z-0607-2014
- Recall status
- Terminated
- CDRH record ID
- 124685
- FDA classified
- 2013-12-29
- Recall event number
- 66979