Z-0607-2018: Synvisc-One (hylan G-F 20), 6mL delivered from a 10mL syringe Product Usage: A elastoviscous high molec

Recall record Z-0607-2018 for Sanofi Genzyme. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0607-2018
Company in FDA record
Sanofi Genzyme
FDA record creation date
Feb 14, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Contamination with Methylbacterium thiocyanatum

Source facts

Action
On December 11, 2017 Sanofi Genzyme distributed URGENT VOLUNTARY DEVICE RECALL notices and reply cards to their customers via courier service. Customers were advised to inspect inventory, immediately discontinue use and return all affected product. On December 19, 2017 Sanofi Genzyme distributed a 2nd notice titled DEVICE RECALL UPDATE NOTICE (2 of 3) notifying customers that the isolated bacterial organism was Methylbacterium thiocyanatum. On December 22, 2017 Sanofi Genzyme distributed a 3rd notice titled DEVICE RECALL UPDATE NOTICE (3 of 3) which included antibiotic sensitivity analysis guidance for the isolated organism. Customers who wish to return product should contact GENCO Pharmaceutical Services at 1-855-838-5782 for a recalled product return kit. Customers are encouraged to complete and return the Business Reply Card via fax to 1-414-459-8791 or via email to recalls@genco.com. Sanofi Genzyme will issue you a credit or replacement product upon receipt of the recalled product by GENCO. All questions of a medical or clinical nature should be directed to Sanofi Medical Information Services in the U.S. at 1-800-633-1610, option 1. Sanofi US Customer service can be reached at 1-800-633-1610 option 7 then 4.
Initiated
2017-12-11
Root cause
Environmental control
Terminated
2023-04-03
Distribution
US Nationwide distribution(36 states)
Recall number
Z-0607-2018
Recall status
Terminated
CDRH record ID
160495
FDA record created
2018-02-14
Recall event number
78792

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗