When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.
Source facts
Action
The firm initiated the recall by letter on 05/24/2018. The letter explained the issue and requested the UPS(s) be disconnected from the system until the system could be repaired.
Initiated
2018-05-24
Root cause
Component change control
Distribution
CA, DC, OH, UT, WI, Canada
Recall number
Z-0607-2019
Recall status
Completed
CDRH record ID
169470
FDA record created
2018-12-13
Recall event number
81680