Otto Bock HealthCare GmbH is recalling Modular Polycentric EBS Knee Joint that were serviced between 27 April 2015 and 17 November 2015 due to a service error where the wrong Loctite was
used during repair.
Source facts
Action
Starting Jan 4 2016, In coordination with the Manufacturer, the US Agent will notify customers via telephone call of the correction on behalf of the Manufacturer. The customers will be requested to return the affected devices back to the service center for correction. The devices will be returned to the customer following correction.
Initiated
2016-01-04
Root cause
Process control
Terminated
2016-05-02
Distribution
Nationwide Distribution.
Recall number
Z-0608-2016
Recall status
Terminated
CDRH record ID
142592
FDA record created
2016-02-01
Recall event number
72960