Reason stated in the source
Device may fail to turn on or unexpectedly attempt to restart, rendering it unable to return to a ready for use state. Additionally, this issue may occur when the device is in standby mode, when attempting to power on to run a self-test. If this occurs, the device will indicate that it is not ready for use.
Source facts
- Action
- Philips Healthcare issued Medical Device Correction Notices dated 10/3/2019 to customers via certified mail. Customers are advised to take the following actions: - The device is safe to use and can remain in service if it does not exhibit any of these behaviors described in this notice. - If you identify a device that exhibit these behaviors, please remove the device from service and contact Philips to request device. - Fill out, sign and send the reply form that is provided as the last page of the letter.
- Initiated
- 2019-10-03
- Root cause
- Process control
- Terminated
- 2023-04-19
- Distribution
- US Nationwide
- Recall number
- Z-0608-2020
- Recall status
- Terminated
- CDRH record ID
- 177814
- FDA record created
- 2019-12-05
- Recall event number
- 84236