Z-0609-2009: DLP(R) Pericardial/Intracardiac Sump Cannulae, 20 Fr., P/N: 12112. This product is intended for draining
Recall record Z-0609-2009 for Medtronic Cardiovascular Revascularization & Surgical Therap. View source facts, labeled dates, and the original FDA record.
Separates: Six complaints were received from two locations. In all cases the sump tip assembly separated from the tubing during device removal and had to be retrieved from the patient. There were no adverse events affecting these patients.
Source facts
Action
Consignees were sent a "Medtronic Urgent Medical Device Recall Notice" dated October 1, 2008. The letter described the product and issue. It also requested its customers to return the product and to complete, sign and fax the Recall Certificate "Medtronic Recall Certificate (FCA 0902)". For further information, please contact Medtronic Cardiovascular Revascularization & Surgical Therap by telephone at 763-391-9000.
Initiated
2008-09-29
Root cause
Nonconforming Material/Component
Terminated
2011-05-07
Distribution
Worldwide Distribution --- including states of NC, PA, IA, WI, IN, VA, IL, KY, and TX, and countries of Australia, France, Hong Kong, Netherlands, and Poland.
Recall number
Z-0609-2009
Recall status
Terminated
CDRH record ID
74901
FDA classified
2009-01-26
Recall event number
50173