Skip to content

Public FDA record

Z-0609-2009: DLP(R) Pericardial/Intracardiac Sump Cannulae, 20 Fr., P/N: 12112. This product is intended for draining

Recall record Z-0609-2009 for Medtronic Cardiovascular Revascularization & Surgical Therap. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0609-2009
Company in FDA record
Medtronic Cardiovascular Revascularization & Surgical Therap
FDA classification date
Jan 26, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Separates: Six complaints were received from two locations. In all cases the sump tip assembly separated from the tubing during device removal and had to be retrieved from the patient. There were no adverse events affecting these patients.

Source facts

Action
Consignees were sent a "Medtronic Urgent Medical Device Recall Notice" dated October 1, 2008. The letter described the product and issue. It also requested its customers to return the product and to complete, sign and fax the Recall Certificate "Medtronic Recall Certificate (FCA 0902)". For further information, please contact Medtronic Cardiovascular Revascularization & Surgical Therap by telephone at 763-391-9000.
Initiated
2008-09-29
Root cause
Nonconforming Material/Component
Terminated
2011-05-07
Distribution
Worldwide Distribution --- including states of NC, PA, IA, WI, IN, VA, IL, KY, and TX, and countries of Australia, France, Hong Kong, Netherlands, and Poland.
Recall number
Z-0609-2009
Recall status
Terminated
CDRH record ID
74901
FDA classified
2009-01-26
Recall event number
50173

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.