There is an anomaly in Medfusion 4000 Syringe Pump Firmware version 1.7.0 that could potentially result in loss of primary audible and visual battery alarm functionality, and interruption of therapy.
Source facts
Action
An Urgent Medical Device recall letter dated 10/28/2019 was sent to customers via email beginning 10/28/2019. The letter identified affected product, stated the issue, and provided instructions for customers to return the pumps. A response form was asked to be returned. Questions can be directed to fieldactions@smiths-medical.com.
Initiated
2019-10-28
Root cause
Software design
Terminated
2024-06-06
Distribution
CO, MD, NY
Recall number
Z-0610-2020
Recall status
Terminated
CDRH record ID
177807
FDA classified
2019-12-12
Recall event number
84232