Z-0612-06: Baxter Auto Syringe AS50 Infusion Pump, product code 1M8550; Baxter Healthcare Corporation, Medication De

Recall record Z-0612-06 for Baxter Healthcare Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0612-06
Company in FDA record
Baxter Healthcare Corp.
FDA classification date
Mar 16, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A non-conforming gasket was used in manufacturing of the specific pumps, causing the pumps to be more susceptible to electro-static discharge, which could increase the potential for the pump to fail during use and alarm.

Source facts

Action
Baxter sent Urgent Device Correction letters dated 2/7/06 to the direct accounts on the same date via first class mail. The letters informed the accounts of the potential for the affected pumps to be susceptible to electro-static discharge, which could increase the potential to fail during use, causing the pump to stop infusing and issue an audible and visual alarm. The accounts were asked to immediately take the pump out of service, contact Baxter at 1-800-843-7867 to arrange for the return of the pumps in order to replace the non-conforming gasket, and complete and return the enclosed reply form confirming receipt of the letter. Any questions were directed to the Center for One Baxter at 1-800-422-9837.
Initiated
2006-02-07
Root cause
Other
Terminated
2008-03-24
Distribution
Nationwide. Arizona, California, Florida, Georgia, Hawaii, Iowa, Indiana, Kentucky, Louisiana, Michigan, Mississippi, North Carolina, Nebraska, New Mexico, New York, Ohio, Oregon, Pennsylvania, Texas, Virginia, Washington, West Virginia
Recall number
Z-0612-06
Recall status
Terminated
CDRH record ID
44132
FDA classified
2006-03-16
Recall event number
34580

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗