Open original FDA record ↗Recall record
- FDA identifier
- Z-0612-06
- Company in FDA record
- Baxter Healthcare Corp.
- FDA classification date
- Mar 16, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A non-conforming gasket was used in manufacturing of the specific pumps, causing the pumps to be more susceptible to electro-static discharge, which could increase the potential for the pump to fail during use and alarm.
Source facts
- Action
- Baxter sent Urgent Device Correction letters dated 2/7/06 to the direct accounts on the same date via first class mail. The letters informed the accounts of the potential for the affected pumps to be susceptible to electro-static discharge, which could increase the potential to fail during use, causing the pump to stop infusing and issue an audible and visual alarm. The accounts were asked to immediately take the pump out of service, contact Baxter at 1-800-843-7867 to arrange for the return of the pumps in order to replace the non-conforming gasket, and complete and return the enclosed reply form confirming receipt of the letter. Any questions were directed to the Center for One Baxter at 1-800-422-9837.
- Initiated
- 2006-02-07
- Root cause
- Other
- Terminated
- 2008-03-24
- Distribution
- Nationwide. Arizona, California, Florida, Georgia, Hawaii, Iowa, Indiana, Kentucky, Louisiana, Michigan, Mississippi, North Carolina, Nebraska, New Mexico, New York, Ohio, Oregon, Pennsylvania, Texas, Virginia, Washington, West Virginia
- Recall number
- Z-0612-06
- Recall status
- Terminated
- CDRH record ID
- 44132
- FDA classified
- 2006-03-16
- Recall event number
- 34580