Centrifuge Rotor Failure: may crack, break, and explode outside the centrifuge.
Source facts
Action
Hettich Instruments notified Distributors by email on 11/21/09, to the Request For Urgent Action notifications, dated 11/23/2009. There is a description of how to identify a recalled rotor as well as a description of the replacement metal rotor that is to be used in place of the list of centrifuges affected by this recall. Customers are to cease using the setup described and contact their distributor or the firm. The firm will replace the plastic housing with a metal housing. Questions should be directed to 978-232-3957.
Initiated
2009-11-21
Root cause
Device Design
Terminated
2011-10-25
Distribution
Worldwide Distribution -- USA, including states of CT and IN, and country of Canada.
Recall number
Z-0612-2010
Recall status
Terminated
CDRH record ID
86894
FDA classified
2010-01-21
Recall event number
53941