It has been determined that, in some cases,
pulley drive components have broken parts or have become loose, mispositioned, or
disconnected from their drive shafts. When these defects occur, the pulley may not turn at all, or
may turn at an incorrect speed, resulting in inadequate washing or mixing of the contents of the
reaction vessel.
Source facts
Action
Beckman Coulter sent an "URGENT PRODUCT CORRECTION" letter dated November 1, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were asked to complete and return an enclosed Response Form. Contact Technical Support for questions regarding this notice.
Initiated
2012-11-01
Root cause
Device Design
Terminated
2013-12-23
Distribution
Worldwide Distribution-USA (nationwide) and the country of Canada.
Recall number
Z-0612-2013
Recall status
Terminated
CDRH record ID
114387
FDA classified
2012-12-22
Recall event number
63632