Z-0612-2019: Stationary Fluoroscopic X-ray System CombiDiagnost R90, Model 709030 Stationary Fluoroscopic X-ray Syste

Recall record Z-0612-2019 for Philips Medical Systems Gmbh, DMC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0612-2019
Company in FDA record
Philips Medical Systems Gmbh, DMC
FDA record creation date
Feb 12, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

kV/mA lockin not functioning as specified which may result in the patient receiving an increased radiation dose.

Source facts

Action
On November 7, 2018, Philips Medical Systems distributed an URGENT MEDICAL DEVICE RECALL" notices dated 10/16/2018 to customers. The following are Actions to Be Taken by the Customer: **If the kV/mA lockin not functioning as specified**: The operator can recognize the issue by observing the increasing mA value. In general the system can be used according to the Instruction for Use without restrictions. **If the Stitching with SkyPlate aborts after first image**: There are no actions to be taken by the user in these situations. The customer is notified at the end of the stitching run that the run failed. Customers can continue to use the device in accordance with this notice and the IfU. Philips will provide an upgrade of all CombiDiagnost R90 systems in the field to Rel.1.0.3 (Eleva package incl. improved clinical EPX database). A Philips Service Engineer will contact you when the Field Action Kit is available to be implemented. Should you need to communicate with Philips with regard to this program, please reference Field Change Order 70900042. If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377.
Initiated
2018-11-07
Root cause
Software design (manufacturing process)
Terminated
2020-09-02
Distribution
Worldwide Distribution: US (nationwide) to states of: AZ, CA, GA, IL, MN, MO, MT, NJ, OH, TX, UT & VA; and to countries of: Australia, Austria, Bangladesh, Belgium, Chile C¿te D'Ivoire, Czech Republic. Egypt, Ethiopia, France, Germany, Ghana, Iran, Italy, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Martinique, Mexico, Nepal, Netherlands, New Zealand, Philippines, Portugal Russian Federation, Rwanda, Saudi Arabia, Serbia, South Africa, Spain, Switzerland, Turkey, United Arab Emirates and United Kingdom.
Recall number
Z-0612-2019
Recall status
Terminated
CDRH record ID
169583
FDA record created
2019-02-12
Recall event number
81706

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗