The femoral impactors have a higher than normal risk of the arm breaking. During impaction the tip of the arm may fracture and separate from the instrument. This may cause a minor delay in surgery. If not retrieved, it may cause an inflammatory response that may require revision surgery.
Source facts
Action
On August 16, 2019, a Urgent Medical device recall for Product Correction" letter was emailed to affected consignees. In addition to informing the consignees about the recall, the
Initiated
2019-08-16
Root cause
Unknown/Undetermined by firm
Terminated
2023-03-02
Distribution
CA
Recall number
Z-0612-2020
Recall status
Terminated
CDRH record ID
177733
FDA classified
2020-01-24
Recall event number
84217