Skip to content

Public FDA record

Z-0613-2010: Haematokrit 24 Centrifuge with 2050 Rotor TYP #2075 and TYP #2075-01. Used for the purpose of determinin

Recall record Z-0613-2010 for Hettich Instruments Management I. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0613-2010
Company in FDA record
Hettich Instruments Management I
FDA classification date
Jan 21, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Centrifuge Rotor Failure: may crack, break, and explode outside the centrifuge.

Source facts

Action
Hettich Instruments notified Distributors by email on 11/21/09, to the Request For Urgent Action notifications, dated 11/23/2009. There is a description of how to identify a recalled rotor as well as a description of the replacement metal rotor that is to be used in place of the list of centrifuges affected by this recall. Customers are to cease using the setup described and contact their distributor or the firm. The firm will replace the plastic housing with a metal housing. Questions should be directed to 978-232-3957.
Initiated
2009-11-21
Root cause
Device Design
Terminated
2011-10-25
Distribution
Worldwide Distribution -- USA, including states of CT and IN, and country of Canada.
Recall number
Z-0613-2010
Recall status
Terminated
CDRH record ID
86895
FDA classified
2010-01-21
Recall event number
53941

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.