Z-0613-2019: AIA-360 Analyzer is capable of performing two methods of immunoassay: an immunoenzymetric (IEMA) or sandw

Recall record Z-0613-2019 for Tosoh Bioscience Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0613-2019
Company in FDA record
Tosoh Bioscience Inc
FDA record creation date
Dec 14, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has become aware of a potential issue with the AIA-360 analyzer that could prevent the table home sensor from detecting the table home position. If the malfunction with the table home sensor occurs on the AIA-360, the sensor board may need to be replaced. If the home sensor fails to detect the home position, the AIA-360 analyzer will cease to function and will be unavailable for use. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received three (3) complaints related to this issue with no serious injuries reported.

Source facts

Action
On October 5, 2018, the firm notified customers of the recall via Urgent Medical Device Recall letter. Customers were informed of the potential issue with the analyzer that could prevent the table home sensor from detecting the table home position. Customers were advised that the table home sensor board will be evaluated and replaced if needed at the next scheduled preventive maintenance visit. Actions to be taken by User: " Thoroughly review the content of this letter. " Continue to use your analyzer. A Tosoh representative will contact you to schedule your next preventive maintenance visit. o If you experience any problems, please contact Tosoh Technical Support for an expedited service visit. " Complete and return the attached Acknowledgement Form. " File this notice with your laboratory records and forward this information to others in your laboratory. If you have obtained unexpected test results, delays in test result reporting , or received any complaints of illness or adverse events associated with the use of the analyzer, please contact Tosoh Technical Support 24 hours a day, seven days a week at (800) 248-6764. Should you have any questions regarding this medical device recall, please feel free to contact Bernadette O'Connell at (800) 248-6764 or by email at bernadette.oconnell@tosoh.com. She will be available to answer any questions Monday- Friday, from 9:00 AM to 5:00 PM (PST).
Initiated
2017-03-22
Root cause
Process control
Terminated
2020-12-09
Distribution
Nationwide distribution. International distribution to Honduras, Caribbean, Chile, Colombia, Costa Rica, Uruguay, Ecuador, Panama, Mexico, Canada, and Venezuela.
Recall number
Z-0613-2019
Recall status
Terminated
CDRH record ID
167620
FDA record created
2018-12-14
Recall event number
81170

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗