Z-0614-2014: Catheter, Continuous Flush The Merit Microcatheter is intended for general intravascular use, including p

Recall record Z-0614-2014 for Merit Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0614-2014
Company in FDA record
Merit Medical Systems, Inc.
FDA classification date
Dec 31, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Merit Medical Systems, Inc. is conducting a recall for 3 lots of Maestro Microcatheters with a graphic discrepancy on the carton/box label end panel. The catheter graphic on the front of the carton box (correct) and the graphic on the end panel (incorrect) do not match.

Source facts

Action
Merit Medical Systems sent an Urgent Product Recall Notice dated December 6, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue use of the Maestro Microcatheters identified and isolate them in a quarantined area until the product is over-labeled by their Merit Sales Representative Customers were instructed to ensure that all personnel to whom the devices were distributed were made aware of the field action. Customers with questions were instructed to contact their sales representative or call 1-801-826-4062. For questions regarding this recall call 801-253-1600.
Initiated
2013-12-06
Root cause
Packaging process control
Terminated
2014-01-27
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada.
Recall number
Z-0614-2014
Recall status
Terminated
CDRH record ID
124756
FDA classified
2013-12-31
Recall event number
67128

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗