There are 3 issues:
1) Under certain conditions, real time waveforms may not be accurately plotted or displayed.
2) The oxygen saturation (SpO2) numeric value displayed on the device may freeze.
3) In auto cycle mode, if the NIBP communication is lost, the pump would not cycle and the non-invasive blood pressure (NIBP) numeric value displayed on the device would not update.
Source facts
Action
The firm initiated the recall on X by letter. The letter explained the issues, potential risk, and provided workarounds for the consignee to use until the new software is ready for release.
Initiated
2018-11-15
Root cause
Software Design Change
Terminated
2024-09-20
Distribution
worldwide
Recall number
Z-0614-2019
Recall status
Terminated
CDRH record ID
168911
FDA record created
2018-12-14
Recall event number
81527