Skip to content

Public FDA record

Z-0614-2021: Carl Zeiss Meditec AG VISUREF150 REF 2227-967 - Product Usage: is a diagnostic instrument used to determi

Recall record Z-0614-2021 for Carl Zeiss Meditec AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0614-2021
Company in FDA record
Carl Zeiss Meditec AG
FDA record creation date
Dec 16, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.

Source facts

Action
On 08/18/2020, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via FedEx overnight informing customers that based on post market surveillance activities, the Recalling Firm found that the communication board on their digital lensmeter and autorefractor were not performing as expected and needs to be exchanged as a precautionary measure. In rare cases the RS232 communication port can fail on the communication board. Both products use the same technology. In the worst-case situation, the safety function of the communication board is insufficient when a connected third-party device is causing an electrical overload at the RS232 interface. This could result in an increased leakage current and electrical shock for the user / patient. If the communication board fails without connecting a third-party device, a burning smell can occur as a result of a thermal event. Customers are to note that the outer housing materials of the lensmeter and autorefractor fulfill the normative requirement concerning flammability. Actions to be taken by the Customer/User: The Recalling Firm is asking customers to disconnect all non-medical grade devices from the affected device RS232 interfaces. If the affected devices are connected to a digital phoropter a disconnection is not needed. Furthermore, the Recalling Firm is asking customers to switch off the affected devices when customers cannot supervise the device, especially when they close their location after working hours. Customers will receive additional information about the handling of this recall from their local service organization. The Recalling Firm is planning to replace the affected component at the next available opportunity. For questions regarding this Recall, contact Customer Support at 877-486-7473.
Initiated
2020-08-06
Root cause
Nonconforming Material/Component
Terminated
2024-12-13
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, MI, MN, NC, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TX, UT, VA, WA and WY. The countries of Argentina, Austria, Australia, Belgium, Bolivia, Canada, Colombia, Croatia, Cyprus, Denmark, France, Germany, Hong Kong, Hungary, India, Ireland, Italy, Japan, Latvia, Luxembourg, Malaysia, Maldives, Netherlands, Portugal, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Turkey, and United Kingdom.
Recall number
Z-0614-2021
Recall status
Terminated
CDRH record ID
183623
FDA record created
2020-12-16
Recall event number
86366

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗