Inaccurate potassium results may be generated. When testing with control fluid and/or patient samples with low or reference range potassium concentration, elevated potassium results may be generated. When testing with samples with a high concentration of potassium, decreased potassium results may be generated
Source facts
Action
Abbott Point of Care issued recall letter on 9/13/18 via FedEx , advising of the problem, recommend action to return all unused cartridges to APOC . If you have any questions regarding this information, please contact Abbott Point of Care Technical Support at 844-256-9531 or your Abbott Point of Care representative.
Initiated
2018-09-13
Root cause
Manufacturing material removal
Terminated
2019-04-04
Distribution
WI
Recall number
Z-0615-2019
Recall status
Terminated
CDRH record ID
167569
FDA record created
2018-12-14
Recall event number
81151