A defect may cause patient image orientation markers to be incorrectly labeled on Multi-Planar Reformation (MPR) images. Left and Right Anatomical orientation markers may be reversed.
Source facts
Action
The firm has issued a notification letter to its consignees on 02/14/2007. This notification letter contains preventive measures.
Initiated
2007-02-14
Root cause
Other
Terminated
2007-12-03
Distribution
Worldwide Distribution ---- USA and countries of Australia, Netherlands, and Canada.
Recall number
Z-0616-2007
Recall status
Terminated