Open original FDA record ↗Recall record
- FDA identifier
- Z-0616-2014
- Company in FDA record
- Steris Corporation
- FDA classification date
- Dec 31, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Customers indicated fluctuations in environmental conditions sites and gradual degradation/deformation of check valve o-rings following frequent contact with S40 sterilant use solution result in nuisance aborts.
Source facts
- Action
- STERIS sent an Urgent Voluntary Field Correction Notice dated November 21, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that a STERIS Field Service Representative would contact their facility to schedule installation of the new hardware and updated software on their SYSTEM IE Liquid Chemical Sterilant Processing System(s). Customers were instructed that they may continue to operate the SYSTEM IE processor prior to the completion of the update. Customers with questions were instructed to contact STERIS Field Service Dispatch at 1-800-333-8828. For questions regarding this recall call 440-392-7519.
- Initiated
- 2013-11-21
- Root cause
- Device Design
- Terminated
- 2014-06-16
- Distribution
- Nationwide Distribution - USA (nationwide) and the District of Columbia and Puerto Rico. Product was also shipped to the following countries: Afghanistan, Bahrain, Hong Kong, Italy, Saudi Arabia, Turkey & United Kingdom.
- Recall number
- Z-0616-2014
- Recall status
- Terminated
- CDRH record ID
- 124471
- FDA classified
- 2013-12-31
- Recall event number
- 67002