- FDA identifier
- Z-0617-03
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Mar 6, 2003
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Unintended yoke rotational motion.
Source facts
- Action
- Siemens sent Customer Advisory letters dated 2/10/03 to all Integrated Orbiter customers. The accounts were informed of the problem with unexpected camera movement in the opposite direction at the beginning of a SPECT study, and were advised to use the emergency stop at any indication of improper camera movement. The customers were advised that they would be contacted by their Siemens Customer Service Representative to schedule a firmware upgrade that will prevent the occurrence of unintended camera rotation. The upgrade will be done free of charge.
- Initiated
- 2003-02-10
- Root cause
- Other
- Terminated
- 2004-04-30
- Distribution
- Nationwide, United Arab Emirates, Argentina, Austria, Azerbaizan,Bangladesh, Belgium, Brazil, Canada, China, Colombia, Cyprus, France, Germany, Great Britain, Greece, Hong Kong, India, Ireland, Italy, Jordan, Japan, Korea, Norway, Pakistan, Romania, Switzerland, Tunisia, Venezuela and the Netherlands.
- Recall number
- Z-0617-03
- Recall status
- Terminated
- CDRH record ID
- 26010
- FDA classified
- 2003-03-06
- Recall event number
- 25544