Reason stated in the source

Unintended yoke rotational motion.

Source facts

Action
Siemens sent Customer Advisory letters dated 2/10/03 to all Integrated Orbiter customers. The accounts were informed of the problem with unexpected camera movement in the opposite direction at the beginning of a SPECT study, and were advised to use the emergency stop at any indication of improper camera movement. The customers were advised that they would be contacted by their Siemens Customer Service Representative to schedule a firmware upgrade that will prevent the occurrence of unintended camera rotation. The upgrade will be done free of charge.
Initiated
2003-02-10
Root cause
Other
Terminated
2004-04-30
Distribution
Nationwide, United Arab Emirates, Argentina, Austria, Azerbaizan,Bangladesh, Belgium, Brazil, Canada, China, Colombia, Cyprus, France, Germany, Great Britain, Greece, Hong Kong, India, Ireland, Italy, Jordan, Japan, Korea, Norway, Pakistan, Romania, Switzerland, Tunisia, Venezuela and the Netherlands.
Recall number
Z-0617-03
Recall status
Terminated
CDRH record ID
26010
FDA classified
2003-03-06
Recall event number
25544

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗