Z-0617-2024: Hookup Accessories used with MEDIVATORS ADVANTAGE PLUS and ADVANTAGE PLUS Pass Thru Automated Endoscope R

Recall record Z-0617-2024 for Steris Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0617-2024
Company in FDA record
Steris Corporation
FDA classification date
Dec 22, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Identification label on affected hookups could become illegible over time due to the label's ink coming off. This may potentially prevent users from being able to effectively identify the hookup and may ultimately result in postponement of patient procedures.

Source facts

Action
On November 13, 2023, the firm notified customers via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to inspect their inventory for affected product. You should not attempt to use a hookup that cannot be effectively identified. Steris will provide either a replacement hookup or replacement label for all affected hookups. If you have questions regarding this matter, please contact Regulatory_Compliance@STERIS.com.
Initiated
2023-11-13
Root cause
Process control
Distribution
Domestic distribution nationwide. Foreign distribution to Australia Canada China Hong Kong Italy Malaysia Netherlands New Zealand Singapore Thailand
Recall number
Z-0617-2024
Recall status
Open, Classified
CDRH record ID
204834
FDA classified
2023-12-22
Recall event number
93566

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗