An RF pulse occurs each time the temperature set point increase button is activated while ablating and this can result in heart rhythm disturbances.
Source facts
Action
An "Important Product Management Information" letter dated November 25, 2002 was sent to the affected consignees (hospitals) and it recommended certain actions to avoid the problem and stated that a software revision, (version 1.36) had been developed to eliminate the problem.
Initiated
2002-11-25
Root cause
Other
Terminated
2003-11-04
Recall number
Z-0618-03
Recall status
Terminated
CDRH record ID
25222
FDA classified
2003-03-06
Recall event number
25111