Open original FDA record ↗Recall record
- FDA identifier
- Z-0618-2009
- Company in FDA record
- ConvaTec
- FDA classification date
- Dec 30, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
It was discovered that a portion of the inner nylon sleeve of the catheter was visible at the balloon end of a sample device.
Source facts
- Action
- Urgent Product Correction Letters were sent by first class mail on November 18, 2008 to Healthcare Professionals and ConvaTec Reps. The reason for the recall was identified along with possible risk to patient. Healthcare professionals were asked to use sufficient lubricating jelly during the insertion process with special care given to those who have a tendency to bleed or are on anti-coagulant medication. Physician should be notified if rectal bleeding or abdominal symptoms develop. ConvaTec reps were asked to contact all health care professionals to whom product or samples were given and reinforce the importance of proper lubrication and following the directions. A return form is to be completed by ConvaTec reps for each visit indicating the name of the Health Care Provider and that the they personally received notice of the product correction. All remaining sample products are to be returned to Besta, Franklin, WI.
- Initiated
- 2008-11-20
- Root cause
- Nonconforming Material/Component
- Terminated
- 2009-03-09
- Distribution
- Worldwide Distribution --- including USA and country of Canada.
- Recall number
- Z-0618-2009
- Recall status
- Terminated
- CDRH record ID
- 75003
- FDA classified
- 2008-12-30
- Recall event number
- 50225