Potential breakage -- product may break while inserting it into the bone during surgery. This may prolong surgery, or, if undetected during surgery, result in retained fragments that may cause metal sensitization or immune response, or necessitate further intervention to retrieve the titanium fragments.
Source facts
Action
Stryker issued an Urgent Device Removal notification letter dated August 21, 2009 to consignees. Consignees were asked to acknowledge receipt and distribute the letter, track reponses, and destroy any returned material and document through weekly summary and records.
Stryker can be contacted at 408 754-2000 about this recall.
Initiated
2009-08-20
Root cause
Material/Component Contamination
Terminated
2010-12-14
Distribution
Nationwide Distribution: to one distributor in NJ.
Recall number
Z-0618-2011
Recall status
Terminated
CDRH record ID
84865
FDA classified
2010-12-14
Recall event number
53182
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.