Open original FDA record ↗Recall record
- FDA identifier
- Z-0619-2011
- Company in FDA record
- Zimmer Inc.
- FDA classification date
- Dec 14, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Mislabeled - Device packages labeled as hooded trial inserts contain standard trial inserts.
Source facts
- Action
- Zimmer contacted the two affected accounts directly by phone on November 15, 2010 to inform them of the recall. Zimmer will monitor the return of the affected instruments and account for the outstanding instruments until the removal action is complete. Each of the affected accounts will be provided with a new trial insert, manufactured and packaged with the correct component. The accounts will be asked to return the affected units back to Zimmer where the returned product will be destroyed. Customers can contact Zimmer at 574 371-8852 concerning this recall.
- Initiated
- 2010-11-15
- Root cause
- Employee error
- Terminated
- 2011-05-23
- Distribution
- Worldwide Distribution: US in the states of KS and UT, and the country of Canada.
- Recall number
- Z-0619-2011
- Recall status
- Terminated
- CDRH record ID
- 95807
- FDA classified
- 2010-12-14
- Recall event number
- 57304