Skip to content

Public FDA record

Z-0620-2009: Foot switch, Part Number 1009.81. Maquet GmbH & Co. KG, 76437 Rastatt, Germany. The foot switches in th

Recall record Z-0620-2009 for MAQUET Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0620-2009
Company in FDA record
MAQUET Inc.
FDA classification date
Jan 15, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A bearing point of the pedal of the foot switch may come loose. With continued use of the foot switch, one or both springs beneath the pedal can fall out of their mounting, and it is possible the pedal comes to rest on the switch causing an unintended motion of the operating table during patient use.

Source facts

Action
An Urgent Device Correction letter (with an enclosed Customer Response form) was sent via Federal Express on Monday, December 8, 2008 to all customers who had received tables. Customers were notified in the letter of the problem and actions to be taken including not to use any affected foot pedal for safety reasons. In addition, customers were instructed to complete and return the Customer Response Form via fax (732-667-1903) to Maquet Inc. Please call USA Maquet, Inc. at toll free number 1-888-627-8383 #1 (then select the option for Surgical Workplace products) for questions.
Initiated
2008-12-08
Root cause
Device Design
Terminated
2012-06-18
Distribution
Worldwide Distribution.
Recall number
Z-0620-2009
Recall status
Terminated
CDRH record ID
75452
FDA classified
2009-01-15
Recall event number
50472

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗