Skip to content

Public FDA record

Z-0620-2013: Companion External Battery, Part Number: 293001-001 The Companion External Battery is a component of the

Recall record Z-0620-2013 for SynCardia Systems Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0620-2013
Company in FDA record
SynCardia Systems Inc.
FDA classification date
Dec 28, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Syncardia Systems is recalling the Companion External Batteries used with the Companion 2 Driver System because they may potentially be at risk of thermal events and/or other physical damage such as melting, charring and/or burning.

Source facts

Action
Syncardia Systems sent a" Urgent Field Safety Notice" dated December 10, 2012, to all affected customers. The letter identified the product the problem and what action the customer needs to take.. The letter informed the customers that there were 3 reported thermal battery events (2 of which occurred on-site at either SynCardia or at the battery manufacturer, Dow Kokam, and one incident that occurred in the field) which caused physical damage to the batteries including burning and/or charring. The batteries were not in use with a patient at time of the reported issue, and there were no reports of any adverse impact to any patient or user. Customers are instructed to follow the directions on the attached Return page, and return their customer response form to SynCardia as soon as possible. Customers with questions are instructed to call (520) 545-1234 or (520) 440-7593.
Initiated
2012-12-05
Root cause
Process control
Terminated
2013-06-07
Distribution
Worldwide Distribution -- US (nationwide)
Recall number
Z-0620-2013
Recall status
Terminated
CDRH record ID
115005
FDA classified
2012-12-28
Recall event number
63915

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗