Open original FDA record ↗Recall record
- FDA identifier
- Z-0620-2013
- Company in FDA record
- SynCardia Systems Inc.
- FDA classification date
- Dec 28, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Syncardia Systems is recalling the Companion External Batteries used with the Companion 2 Driver System because they may potentially be at risk of thermal events and/or other physical damage such as melting, charring and/or burning.
Source facts
- Action
- Syncardia Systems sent a" Urgent Field Safety Notice" dated December 10, 2012, to all affected customers. The letter identified the product the problem and what action the customer needs to take.. The letter informed the customers that there were 3 reported thermal battery events (2 of which occurred on-site at either SynCardia or at the battery manufacturer, Dow Kokam, and one incident that occurred in the field) which caused physical damage to the batteries including burning and/or charring. The batteries were not in use with a patient at time of the reported issue, and there were no reports of any adverse impact to any patient or user. Customers are instructed to follow the directions on the attached Return page, and return their customer response form to SynCardia as soon as possible. Customers with questions are instructed to call (520) 545-1234 or (520) 440-7593.
- Initiated
- 2012-12-05
- Root cause
- Process control
- Terminated
- 2013-06-07
- Distribution
- Worldwide Distribution -- US (nationwide)
- Recall number
- Z-0620-2013
- Recall status
- Terminated
- CDRH record ID
- 115005
- FDA classified
- 2012-12-28
- Recall event number
- 63915