Open original FDA record ↗Recall record
- FDA identifier
- Z-0620-2014
- Company in FDA record
- GE Healthcare It
- FDA classification date
- Jan 2, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
GE Healthcare is aware of a potential safety issue with the use of GE Centricity Laboratory Instrument Interface (IF) where free-text sent from IM is not being transcribed into free-text in iNET. The instrument may upload results to IF/iNET when order/request is not yet in IF/iNET. In such a situation, test level supporting free-text sent from the instrument is discarded (not captured).
Source facts
- Action
- GE Healthcare sent an URGENT MEDICAL DEVICE CORRECTION letter dated March 12, 2013 to all affected customers. The letter notified customers of the problem, suggested that customers experiencing this problem should retransmit the result from the instrument once the order exists in iNET and verify that the item level free-text is present in iNET, and to contact their GE service representative to request or install the updated software. Customers with additional questions can contact the GE Healthcare IT Help Desk at 1-847-589-8496 or tgg.nahelpdesk@med.ge.com. For questions regarding this recall call 847-589-8496.
- Initiated
- 2013-03-12
- Root cause
- Device Design
- Terminated
- 2016-06-10
- Distribution
- Worldwide Distribution - US including MN and Internationally to India, Australia, Canada, and England.
- Recall number
- Z-0620-2014
- Recall status
- Terminated
- CDRH record ID
- 117106
- FDA classified
- 2014-01-02
- Recall event number
- 64807