Open original FDA record ↗Recall record
- FDA identifier
- Z-0620-2025
- Company in FDA record
- Olympus Corporation of the Americas
- FDA classification date
- Dec 5, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
High-level disinfection ("HLD") with Acecide-C disinfectant in the Olympus OER-Pro Automated Endoscope Reprocessor ("AER") has been removed as a compatible reprocessing method from the Instructions for Use when reprocessing the GIF-1THI90 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE ( GIF-1TH190 ).
Source facts
- Action
- Olympus notified consignees via letter on about 10/31/2024. Consignees were informed of the issue and of the upcoming revision of the labeling such that the GIF-1TH190 is no longer a compatible endoscope for reprocessing with Acecide-C in the OER-Pro. They were instructed that they would be contacted by their Olympus sales representative within the month to discuss potential OER-Elite and GIF-1TH190 transition programs. Additionally, they were instructed to carefully read the notification, ensure reprocessing staff and other users of the OER-Pro AER are aware of the corrective action, acknowledge receipt of the notification through the Olympus web portal, and to identify and notify recipients of the affected devices if they were further distributed.
- Initiated
- 2024-10-31
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.
- Recall number
- Z-0620-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210864
- FDA classified
- 2024-12-05
- Recall event number
- 95641