- FDA identifier
- Z-0621-2024
- Company in FDA record
- Atrium Medical Corporation
- FDA classification date
- Dec 28, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A sterilization nonconformance prematurely aged the product by 48.3 days, resulting in the expiration date on the label not being accurate. This nonconforming product was inadvertently distributed to customers. Increased risk to patients and healthcare providers using the product after the shortened expiration date of November 17, 2025.
Source facts
- Action
- On November 1, 2023, the firm issued URGENT MEDICAL DEVICE - Removal letters to affected customers. Customers were instructed to remove any product from the affected lots from areas of use. Customers with affected product are entitled to a replacement at no cost to their facility. Customers will receive replacements upon their acknowledgement that they have affected product for return. Distributors should forward recall documentation to affected customers. Customers with questions may contact their Atrium/Getinge representative or call the Atrium/Getinge Customer Support at (603) 880-1433 (press option 1), Monday through Friday, 9AM-6PM Eastern Time Zone.
- Initiated
- 2023-11-01
- Root cause
- Nonconforming Material/Component
- Distribution
- Domestic distribution nationwide. No foreign distribution.
- Recall number
- Z-0621-2024
- Recall status
- Open, Classified
- CDRH record ID
- 204714
- FDA classified
- 2023-12-28
- Recall event number
- 93553