Z-0622-2022: bellavista 1000 ventilator, Catalog #301.100.030 and bellavista 1000e ventilator, 301.100.130. The bella

Recall record Z-0622-2022 for Vyaire Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0622-2022
Company in FDA record
Vyaire Medical
FDA classification date
Feb 16, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential cessation in ventilation can occur under specific conditions.

Source facts

Action
International consignees were notified of the correction via email notification beginning 12/20/2021 containing a Field Safety Notice, an Appendix to the Field Safety Notice which lists the affected units, an FSCA Distributor Response Form and an FSCA End User Response form, all in English, with the notation that localized versions were currently in translation and will be forwarded as soon as available, if applicable. The email provided information on the affected model numbers and actions to be taken by the distributor. These actions included notifying all affected end users by providing them with the FSCA package containing the notice and the list of affected units, notify the appropriate regulatory authorities as applicable, return the completed and signed FSCA Distributor Response form, and letting them know that if any user facilities have distributed affected products to other persons or facilities, promptly forward a copy of the Field Safety Notice and the list of affected units. The email informs them that imtmedical ag will inform them of the availability of the software patch in a separate communication for installation. The recalling firm, Vyaire Medical, issued two different notifications titled "URGENT: MEDICAL DEVICE CORRECTION" to distributors and end users on 12/23/2021 via electronic communication. Physical copies of the notification will be issued to each consignee via UPS Next Day Air delivery. The Distributor letter informed the distributor of the issue and provided a problem description, problem root cause, and said that in order to prevent adverse health consequences associated with this failure mode during use of the bellavista 1000 and bellavista 1000e ventilators, hardware generation G6, end users are encouraged to follow the User Manual and take the immediate mitigation actions listed in the End-User Notification package. The actions to be taken by the distributor included: (1) Immediately notify all affected end users by providi
Initiated
2021-12-20
Root cause
Software change control
Terminated
2024-08-28
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, CA, CO, CT, DC, FL, GA, ID, IL, IN, LA, MA, MD, MI, MO, MS, NC, NJ, NY, OH, OK, PA, SD, TX, VA, WA, WI and WV, including Puerto Rico and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Dominican Republic, Egypt, El Salvador, France, Germany, Guatemala, India, Indonesia, Iran, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lithuania, Malaysia, Mexico, Namibia, Netherlands, Nicaragua, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, South Africa, South Korea, Spain, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam, and Zimbabwe.
Recall number
Z-0622-2022
Recall status
Terminated
CDRH record ID
191357
FDA classified
2022-02-16
Recall event number
89366

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗