Skip to content

Public FDA record

Z-0623-03: e.cam Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Sieme

Recall record Z-0623-03 for Siemens Medical Solutions USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0623-03
Company in FDA record
Siemens Medical Solutions USA, Inc.
FDA classification date
Mar 6, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Unexpected release of the radial and rotate brakes in the case of multiple simultaneous fault conditions.

Source facts

Action
Siemens sent Customer Advisory letters dated 2/10/03 to all e.cam customers. The accounts were informed of the problem with unexpected camera movement when multiple fault cocnditions are presnet, causing the brake system to release and the roation of the detector assemblies, and were advised to use the emergency stop at any indication of improper camera movement. The customers were advised that they would be contacted by their Siemens Customer Service Representative to schedule a firmware upgrade that will prevent the occurrence of unintended release of the radial and rotate brakes. The upgrade will be done free of charge.
Initiated
2003-02-10
Root cause
Other
Terminated
2004-09-15
Distribution
Nationwide, United Arab Emirates, Argentina, Austria, Australia, Bosnia, Barbados, Bangladesh, Belgium, Bolivia, Brazil, Camaroon, Canada, Chile, China, Colombia, Republic of Congo, Czech Republic, Denmark, Dominican Republic, El Salvador, Equador, Egypt, France, Finland, Germany, Great Britain, Georgia, Greece, Hong Kong, Honduras, Ireland, Israel, India, Italy, Iraq, Jordan, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Mexico, Morocco, Madagascar, Myanmar, Mauritius, Malaysia, Niger, Norway, New Zealand, Oman, Peru, Philippines, Pakistan, Poland, Potugal, Russia, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine, Uzbekistan, South Africa and the Netherlands.
Recall number
Z-0623-03
Recall status
Terminated
CDRH record ID
26012
FDA classified
2003-03-06
Recall event number
25546

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.