Open original FDA record ↗Recall record
- FDA identifier
- Z-0623-03
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Mar 6, 2003
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Unexpected release of the radial and rotate brakes in the case of multiple simultaneous fault conditions.
Source facts
- Action
- Siemens sent Customer Advisory letters dated 2/10/03 to all e.cam customers. The accounts were informed of the problem with unexpected camera movement when multiple fault cocnditions are presnet, causing the brake system to release and the roation of the detector assemblies, and were advised to use the emergency stop at any indication of improper camera movement. The customers were advised that they would be contacted by their Siemens Customer Service Representative to schedule a firmware upgrade that will prevent the occurrence of unintended release of the radial and rotate brakes. The upgrade will be done free of charge.
- Initiated
- 2003-02-10
- Root cause
- Other
- Terminated
- 2004-09-15
- Distribution
- Nationwide, United Arab Emirates, Argentina, Austria, Australia, Bosnia, Barbados, Bangladesh, Belgium, Bolivia, Brazil, Camaroon, Canada, Chile, China, Colombia, Republic of Congo, Czech Republic, Denmark, Dominican Republic, El Salvador, Equador, Egypt, France, Finland, Germany, Great Britain, Georgia, Greece, Hong Kong, Honduras, Ireland, Israel, India, Italy, Iraq, Jordan, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Mexico, Morocco, Madagascar, Myanmar, Mauritius, Malaysia, Niger, Norway, New Zealand, Oman, Peru, Philippines, Pakistan, Poland, Potugal, Russia, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine, Uzbekistan, South Africa and the Netherlands.
- Recall number
- Z-0623-03
- Recall status
- Terminated
- CDRH record ID
- 26012
- FDA classified
- 2003-03-06
- Recall event number
- 25546