Philips Medical Systems North America Co. Phillips
FDA classification date
Sep 25, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
System fails to meet requirements when it is used at kV levels greater than 115 kV at certain frame rates and in certain acquisition modes, as grid switch leakage may occur.
Source facts
Action
A Customer Notification letter describing the failure, the means, by which a customer can prevent the failure mode from occurring, and actions planned by Philips to correct the problem were sent to all users of the affected systems. Philips will modify the system software and will be installed at no cost to the customer.
Initiated
2008-12-30
Root cause
Radiation Control for Health and Safety Act
Terminated
2010-09-28
Distribution
Nationwide distribution including states CA, CO, IA, ID, KS, MN, NC, NJ, NY, OH, SC, TX, UT, VA, and WI.
Recall number
Z-0623-2009
Recall status
Terminated
CDRH record ID
76931
FDA classified
2009-09-25
Recall event number
50748