Z-0623-2013: Space Pump Pole Clamps for the Infusomat Space Pump, model number 8713130. The Pole Clamp is an accessor

Recall record Z-0623-2013 for B. Braun Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0623-2013
Company in FDA record
B. Braun Medical, Inc.
FDA classification date
Dec 31, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Firm has become aware of the potential for breakage of the pole clamp locking grids, or rasters, in the Space Pole Clamps.

Source facts

Action
B. Braun Medical, Inc sent an Urgent Medical Device letter dated March 23, 2012 , to all affected customers. The letter identified the product the problem and the action needed by the customer to take. B. Braun has implemented an improvement to the pump locking mechanism of t he older style pole clamps and will be upgrading any older style pole clamps that may remain within your facility. Please return the completed form from attachment 1 in the self-addressed envelope provided or fax the form to (610) 849-1197 within two (2) weeks of receipt\. If you require any additional information or have any questions, comments or concerns please contact your B. Braun Medical Customer Service at 800-627-7867.
Initiated
2012-03-23
Root cause
Device Design
Terminated
2015-02-02
Distribution
Worldwide Distribution--US (nationwide) including the states of AL, AZ, NC and TN.
Recall number
Z-0623-2013
Recall status
Terminated
CDRH record ID
114927
FDA classified
2012-12-31
Recall event number
63862

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗