Skip to content

Public FDA record

Z-0625-2016: Zimmer, Virage OCT Spinal Fixation System, Item Number: 07.01738.001, a posterior system intended to ach

Recall record Z-0625-2016 for Zimmer Biomet, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0625-2016
Company in FDA record
Zimmer Biomet, Inc.
FDA record creation date
Jan 9, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Zimmer Biomet is initiating a recall of the Virage Occipital Eyelet, as it was found through product testing that the staking of the set screw was out of specification and can interfere with the insertion of the occipital rod when the set screw is staked too low.

Source facts

Action
Consignees were sent a Zimmer Biomet "Urgent Medical Device Recall" letter dated November 17, 2015. One letter was addressed to Risk Manager and Surgeons and a separate letter was sent to Distributors, Sales Representatives, and Distributor Operations Mangers. The letters described the problem and the product involved in the recall. The letter described the "Risk" and the responsibilities of consignees and requested consignees to quarantine the product. Risk Managers and Surgeons were requested to complete and return the Certificate of Acknowledgment form. Distributors, Sales Representatives, and Distributor Operations Mangers were requested to complete and return the Inventory Return certification Form. For questions consignees can call the customer call center at 1-866-774-6368 between 8:00 am and 5:00 pm EST.
Initiated
2015-11-17
Root cause
Process control
Terminated
2016-08-12
Distribution
US: AL, AZ, CA, CO, FL, GA, ID, MI, MN, MO, NV, NJ, OK, OR, PA, SD, TN, TX, WA, WI. OUS: AUSTRALIA, GERMANY, SWITZERLAND, SWEDEN.
Recall number
Z-0625-2016
Recall status
Terminated
CDRH record ID
142323
FDA record created
2016-01-09
Recall event number
72832

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗