Open original FDA record ↗Recall record
- FDA identifier
- Z-0626-2014
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA classification date
- Jan 3, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Positive bias on the lots listed compared to alternate methods and a higher frequency of "Above Assay Range" flags with the lots listed.
Source facts
- Action
- Siemens sent an Urgent Medical Device Recall letter in April 2013 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discard any remaining inventory of the affected product. Siemens will replace any unused inventory at no charge. Customers were asked to complete the attached form and fax it to 302-631-8467 to indicate they have received the information. Customers were asked to retain the letter with their laboratory records and forward the letter to those who may have received the affected product. Customers with questions were instructed to contact their Siemens Technical Solutions Center or their local Siemens technical support representative. For questions regarding this recall call 1-800-441-9250.
- Initiated
- 2013-04-22
- Root cause
- Process control
- Terminated
- 2014-02-12
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Canada and Mexico.
- Recall number
- Z-0626-2014
- Recall status
- Terminated
- CDRH record ID
- 123079
- FDA classified
- 2014-01-03
- Recall event number
- 65409