Open original FDA record ↗Recall record
- FDA identifier
- Z-0626-2019
- Company in FDA record
- Covidien LLC
- FDA record creation date
- Dec 18, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
Source facts
- Action
- On November 15, 2018 Medtronic issued UPDATED URGENT MEDICAL DEVICE RECALL notices to customers. Customers are advised to take the following actions: 1. Please immediately quarantine and discontinue use of the affected products. 2. Return affected product as indicated. 3. If you have distributed the products, please promptly forward the information from the letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory. If you have any questions or concerns, please do not hesitate to contact your Medtronic representative or Technical Support at (800) 448-3644, Option 3, Option 8.
- Initiated
- 2018-11-15
- Root cause
- Nonconforming Material/Component
- Distribution
- US Nationwide
- Recall number
- Z-0626-2019
- Recall status
- Open, Classified
- CDRH record ID
- 169030
- FDA record created
- 2018-12-18
- Recall event number
- 81595