Open original FDA record ↗Recall record
- FDA identifier
- Z-0628-2011
- Company in FDA record
- Advanced Sterilization Products
- FDA classification date
- Dec 16, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The recall was initiated because Advanced Sterilization Products (ASP) received three customer reports of elevated temperature of the high-level disinfectant solution used in the ASP AER as a result of a possible temperature monitoring system malfunction. This recall communication advised consignees that if high-level disinfectants are exposed to elevated temperatures, fumes may occur and resul
Source facts
- Action
- Advanced Sterilization Products sent an URGENT: PRODUCT CORRECTION letter dated February 19, 2010, to all customers. The letter identified the product, the problem, and the action to be taken by the customer. The recall communication was initiated with ASP forwarding a "Urgent: Product Correction" letter with attached customer fax-back response form to the affected customers. Customers were also instructed to complete and sign the enclosed customer fax-back response form indicating that they have disconnected the heater on the ASP AER according to the instructions provided and fax the form to 1-888-345-1435. If customers have any questions or need assistance, they were instructed to call ASP Professional Services at 888-783-7723, option 3 then option 1.
- Initiated
- 2010-02-19
- Root cause
- Pending
- Terminated
- 2013-07-30
- Distribution
- Worldwide Distribution -- United States, Argentina Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, Hungary, Mexico, Panama, Guatemala, Honduras, Poland, Puerto Rico, Virgin Islands, Singapore, Taiwan, Venezuela, Croatia, and Bulgaria.
- Recall number
- Z-0628-2011
- Recall status
- Terminated
- CDRH record ID
- 95737
- FDA classified
- 2010-12-16
- Recall event number
- 57281