Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. The affected beds may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes.
Source facts
Action
Stryker sent an Urgent Medical Device Correction letterdated December 18, 2012, to all affected customer. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to locate the affected product and ensure that the brakes are holding properly. Consignees were directed to contact Stryker (1-800-327-0770, option 2) to make arrangements to replace all brake rings and continue to routinely inspect brake functionality. Customers were asked to Fax (269-488-8691) the enclosed acknowledgment and Stryker field service will contact customers to schedule service. Consignees were directed to forward a
copy of this notice to the new users and advise us of their new location.
For questions regarding this recall call 269-488-8691.
Initiated
2012-12-18
Root cause
Process control
Terminated
2013-08-30
Distribution
Worldwide Distribution - Nationwide and Canada
Recall number
Z-0628-2013
Recall status
Terminated
CDRH record ID
114351
FDA classified
2013-01-03
Recall event number
63602