Open original FDA record ↗Recall record
- FDA identifier
- Z-0629-2009
- Company in FDA record
- Accuray Inc
- FDA classification date
- Jan 17, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
System may use random incorrect data to calculate dose. Resulting dose calculation error can exceed 100% of correct dose which may lead to serious patient injury.
Source facts
- Action
- An Urgent Advisory Notification letter dated November 14, 2008 was issued to all sites via Fed Ex. The letter identifies the issue and required actions. The user was instructed not to modify current extrapolation depth settings and to contact Accuray Incorporated Customer Support prior to making any changes. Please direct any questions or concerns regarding this issue to Accuray Incorporated Customer Support at 1-877-668-8667 (USA) or 1-408-716-4700 (outside of USA) or customersupport@accuray.com.
- Initiated
- 2008-11-14
- Root cause
- Software design
- Terminated
- 2010-12-02
- Distribution
- Nationwide Distribution. Product was distributed to 4 medical facilities in the US.
- Recall number
- Z-0629-2009
- Recall status
- Terminated
- CDRH record ID
- 74985
- FDA classified
- 2009-01-17
- Recall event number
- 50207