Skip to content

Public FDA record

Z-0629-2013: Medical Surgical Bed, Model 3005S3EX Medical Surgical Bed The S3" MedSurg Bed, Model 3005 is intended to

Recall record Z-0629-2013 for Stryker Medical Division of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0629-2013
Company in FDA record
Stryker Medical Division of Stryker Corporation
FDA classification date
Jan 3, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes.

Source facts

Action
Stryker sent an Urgent Medical Device Correction letterdated December 18, 2012, to all affected customer. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to locate the affected product and ensure that the brakes are holding properly. Consignees were directed to contact Stryker (1-800-327-0770, option 2) to make arrangements to replace all brake rings and continue to routinely inspect brake functionality. Customers were asked to Fax (269-488-8691) the enclosed acknowledgment and Stryker field service will contact customers to schedule service. Consignees were directed to forward a copy of this notice to the new users and advise us of their new location. For questions regarding this recall call 269-488-8691.
Initiated
2012-12-18
Root cause
Process control
Terminated
2013-08-30
Distribution
Worldwide Distribution - Nationwide and Canada
Recall number
Z-0629-2013
Recall status
Terminated
CDRH record ID
114352
FDA classified
2013-01-03
Recall event number
63602

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗