Open original FDA record ↗Recall record
- FDA identifier
- Z-0629-2014
- Company in FDA record
- Fresenius Medical Care Holdings, Inc.
- FDA classification date
- Jan 6, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
2008 Series Hemodialysis Machines complaints of saline bags inappropriately filling during recirculation and setup
Source facts
- Action
- Fresenius Medical issued a Product Notification letter dated November 12, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers who encounter a saline bag that is inappropriately filling, should discontinue the setup and remove the machine from service until the drain line and drain-associated hydraulic components can be checed by a qualified service technician. Customers were also instructed to ensure that the hemodialysis machine is installed and operated in compliance with the Machine Specification section of their operator's manual, including the specification located in the Water/Drain Section. Customers with questions were instructed to call 1-800-227-2572. For questions regarding this recall call 800-662-1237.
- Initiated
- 2013-11-12
- Root cause
- Under Investigation by firm
- Terminated
- 2019-03-26
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally Mexico and Canada.
- Recall number
- Z-0629-2014
- Recall status
- Terminated
- CDRH record ID
- 122784
- FDA classified
- 2014-01-06
- Recall event number
- 66622