Z-0629-2018: Hemotherm Dual Reservoir Cooler/Heater System Models 400CE and 400MR Product Usage: A dual reservoir, wat

Recall record Z-0629-2018 for Cincinnati Sub-Zero Products LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0629-2018
Company in FDA record
Cincinnati Sub-Zero Products LLC
FDA classification date
Feb 13, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated Instructions For Use includes a more robust cleaning & disinfection procedure.

Source facts

Action
Between November 27, 2017 - December 8, 2017, Cincinnati Sub-Zero Products LLC, distributed Urgent Medical Device Correction notices to their customers via Email and USPS. This Medical Device Correction has been initiated to provide a labeling update to the cleaning & disinfection procedures. The update addresses the disinfection process, cleaning of reusable accessories, as well as clarifying the water type that should be used with the Hemotherm. Customers are encouraged to do the following: 1) Discard your current Hemotherm Operation & Technical Manual and/or Operation Manual. 2) Replace your current Hemotherm Operation & Technical Manual and/or Operation Manual with the enclosed updated Manual(s). For your convenience, all updates have been highlighted in the manuals. " For Hemotherm 400CE users, utilize the P/N 56418-R Operation & Technical Manual and P/N 56075-G Operation Manuals. " For Hemotherm 400MR users, utilize the P/N 56234-S Operation Manual. All customers are encouraged to complete and return the enclosed response form by fax or email to FA2016-009@genthermcsz.com; Fax: 513-772-9119. Customers with questions or concerns, please contact the Field Action Coordinator: Kathy DeSmidt, E-mail: FA2016-009@genthermcsz.com, Phone: 513-772-8810 or 800-989-7373.
Initiated
2017-11-27
Root cause
Device Design
Terminated
2018-07-02
Distribution
US Nationwide Distribution
Recall number
Z-0629-2018
Recall status
Terminated
CDRH record ID
158835
FDA classified
2018-02-13
Recall event number
78173

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗