The Color Cuff Sterile Disposable Tourniquet Cuffs with Quick-Connect Fitting may have an incorrect end-connector.
A customer reported having received a luer connector instead of a quick connect fitting. An incorrect connector may
lead to an inoperable system due to incompatibility with the Smart Pump.
Source facts
Action
Distributor was notified by E-mail 11/26/2013, with follow-up UPS letters sent to Hospital Risk/ Materials Managers identifying the product, reason for the recall, and risk. The notice asked consignees to find and quarantine all implicated product, with hospitals requested to complete the Business Reply Form and FAX to 866-521-2762. Stryker will email a pre-paid shipper and credit all accounts. Question and concerns should be addressed to Kelly Jo Whipple 269-389-2921 or kellyjo.whipple@stryker.com
Initiated
2013-12-06
Root cause
Component change control
Terminated
2014-02-27
Distribution
US Distribution: including states of: MI, OH, and TX.
Recall number
Z-0630-2014
Recall status
Terminated
CDRH record ID
124590
FDA classified
2014-01-06
Recall event number
67040