Z-0630-2022: BD Veritor Plus Analyzer - intended to provide rapid test results in near-patient settings. Catalog Numb

Recall record Z-0630-2022 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0630-2022
Company in FDA record
Becton Dickinson & Co.
FDA classification date
Feb 9, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

May have the potential to overheat and/or cause fire, the issue can occur when an unauthorized alternative AC power adapter, and not the one provided by BD is used to power the instrument.

Source facts

Action
***UPDATE June 6, 2022**** BD issued a Medical Device Product Advisory letter to Distributors and Users dated July 6, 2022 as a follow-up notification to the voluntary product advisory dated January 5, 2022. In light of this recent complaint, BD is providing additional recommendations to customers to avoid a fire until BD can replace affected units in the field. Letter states: 1. Ensure the contents of this Product Advisory are read and understood by those within your organization. 2. To verify if your BD Veritor" Plus Analyzer(s) is impacted and to complete the acknowledgment distributor response form, you can visit the following website https://go.bd.com/Veritor-Analyzer-Notice.html and enter your serial number(s). 3. AT ALL TIMES USE ONLY the 5V 3A, 15W AC/DC power adapter provided with the BD Veritor" Plus Analyzers. If needed, a replacement AC/DC power adapter (catalog number 445201) can be purchased by calling BD Customer Service. 4. Affix the label(s) provided with this notification to your BD provided AC/DC power adapter(s) to ensure the correct power adapter is always connected to the analyzer(s). BD will contact your account to coordinate replacement of your affected analyzer(s) with expiration dates between December 2021 and August 2024. In the meantime, please implement the recommendations detailed above to mitigate risk of fire. 2. BD will provide credit or replacement for any affected analyzers that are disposed of in response to this notification. ------------------------------------------------------------------------------------------------------------------ BD issued Medical Device Advisory letter dated 1/05/22 to Lab Directors/Risk Managers and Distributors . Letter states reason for recall, health risk and action to take: 1. Ensure the contents of this Product Advisory are read and understood by those within your organization. 2. USE ONLY the AC power adapter provided with the BD Veritor" Plus Analyzers. An image of the AC power adapter provide
Initiated
2022-01-05
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Armenia, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Dominican Republic, Ecuador, Fiji, Guatemala, Hong Kong, Honduras, Haiti, Indonesia, India, Jamaica, Cambodia, South Korea, Macao, Mexico, New Zealand, Panama, Peru, Philippines, Singapore, El Salvador, Thailand, Trinidad and Tobago, Taiwan, Uruguay, Vietnam.
Recall number
Z-0630-2022
Recall status
Open, Classified
CDRH record ID
191328
FDA classified
2022-02-09
Recall event number
89358

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗