Skip to content

Public FDA record

Z-0630-2024: KeySplint Hard Clear - Indicated for the fabrication of orthodontic and dental appliances such as mouthgu

Recall record Z-0630-2024 for Keystone Industries. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0630-2024
Company in FDA record
Keystone Industries
FDA classification date
Jan 3, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Keyprint Keysplint Hard Clear 1 kg resin bottles were inadvertently packaged with the wrong Instructions for Use. Keysplint Hard Clear were placed inside a zip lock bag along with KeySplint Soft Instructions for Use (IFU).

Source facts

Action
Keystone Industries issued Urgent Medical Device Recall/Correction letter to Distributors between 10/20/2023 and 10/23/2023. Letter states reason for recall, health risk and action to take: For immediate access, the correct IFU is available following this link: 4916111-NATLAB-1372-KeySplint_Hard_DOMESTIC_IFU_WEB.pdf (keystoneindustries.com) IMMEDIATE ACTIONS TO BE TAKEN 1. Review product inventory and quarantine all affected product on hand. Open zip lock bag and remove the incorrect IFU. Place an X across the front panel of the IFU using a marker or pen. Take a picture of the IFU with an X then discard the IFUs. Record the information on the attached acknowledgement form. 2. Keystone Industries will send you the correct IFUs for any items remaining in your inventory upon receipt of the acknowledgement form. 3. Review distribution records and identify all customers who were shipped the affected product lot. 4. Send End User Notifications provided to you with this notification to all identified customers that received product. Record evidence of notification on the acknowledgement form. Complete the attached Acknowledgement of Medical Device Recall Form and return within 5 days of receiving this notice. Email the completed form to amcclure@keystoneind.com If you have any questions after reviewing this notice, please email using the address below or call Caleb Barylski at 856-548-5205 between 8:00 am and 5:00 pm EST, Monday through Friday.
Initiated
2023-11-20
Root cause
Labeling mix-ups
Distribution
Worldwide distribution - US Nationwide and the country of China.
Recall number
Z-0630-2024
Recall status
Open, Classified
CDRH record ID
204643
FDA classified
2024-01-03
Recall event number
93522

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗