Open original FDA record ↗Recall record
- FDA identifier
- Z-0630-2024
- Company in FDA record
- Keystone Industries
- FDA classification date
- Jan 3, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Keyprint Keysplint Hard Clear 1 kg resin bottles were inadvertently packaged with the wrong Instructions for Use. Keysplint Hard Clear were placed inside a zip lock bag along with KeySplint Soft Instructions for Use (IFU).
Source facts
- Action
- Keystone Industries issued Urgent Medical Device Recall/Correction letter to Distributors between 10/20/2023 and 10/23/2023. Letter states reason for recall, health risk and action to take: For immediate access, the correct IFU is available following this link: 4916111-NATLAB-1372-KeySplint_Hard_DOMESTIC_IFU_WEB.pdf (keystoneindustries.com) IMMEDIATE ACTIONS TO BE TAKEN 1. Review product inventory and quarantine all affected product on hand. Open zip lock bag and remove the incorrect IFU. Place an X across the front panel of the IFU using a marker or pen. Take a picture of the IFU with an X then discard the IFUs. Record the information on the attached acknowledgement form. 2. Keystone Industries will send you the correct IFUs for any items remaining in your inventory upon receipt of the acknowledgement form. 3. Review distribution records and identify all customers who were shipped the affected product lot. 4. Send End User Notifications provided to you with this notification to all identified customers that received product. Record evidence of notification on the acknowledgement form. Complete the attached Acknowledgement of Medical Device Recall Form and return within 5 days of receiving this notice. Email the completed form to amcclure@keystoneind.com If you have any questions after reviewing this notice, please email using the address below or call Caleb Barylski at 856-548-5205 between 8:00 am and 5:00 pm EST, Monday through Friday.
- Initiated
- 2023-11-20
- Root cause
- Labeling mix-ups
- Distribution
- Worldwide distribution - US Nationwide and the country of China.
- Recall number
- Z-0630-2024
- Recall status
- Open, Classified
- CDRH record ID
- 204643
- FDA classified
- 2024-01-03
- Recall event number
- 93522