Horiba Medical is recalling the ABX PENTRA AST CP because during the AST application validation on the Pentra C200 using an Audit Systems linearity kit, it was determined that two higher values for AST (expected at 800 U/L and 1000 U/L) were recorded at 280 and 199 U/L without alarm.
Source facts
Action
Horiba Medical sent a Field Safety Notification and Acknowledgement Letter dated November 20, 2012, to all affected customers. The letter notified the customers of the problems identified and the actions to be taken. Customers were instructed to complete and return the enclosed response form within 10 calendar days via mail, email or fax. Customers with questions were instructed to contact their local representative.
For questions regarding this recall call 714-273-9254.
Initiated
2012-10-10
Root cause
Software design
Terminated
2013-09-12
Distribution
Nationwide Distribution including CA, OK, NY, FL, AZ, NY, VA, and TX.
Recall number
Z-0632-2013
Recall status
Terminated
CDRH record ID
114864
FDA classified
2013-01-03
Recall event number
63838